Hydrogen ion-selective poly(vinyl chloride) membrane electrode based on a p-tert-butylcalix[4]arene-oxacrown-4

Yükleniyor...
Küçük Resim

Tarih

2004

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

ELSEVIER SCIENCE BV

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

A hydrogen ion-selective poly(vinyl chloride) (PVC) membrane electrode was developed using p-tert-butylcalix[4]arene-oxacrown-4 as ionophore. Apart from the ionophore, plasticisers and lipophilic anions were blended, in various proportions, with PVC in tetrahydrofurane and n-fixtures so prepared were poured onto glass surfaces to form the membrane. 2-Nitrophenylpentylether has proved to be the best alternative as plasticiser and the lipophilic anions tried have turned out to have an adverse effect on the pH response. The electrode of the optimum characteristics had a composition of 2% p-tert-butylcalix[4]arene-oxacrown-4, 68.3% o-NPPE and 29.7% PVC. The electrode showed an apparent Nernstian response in the pH range 2-11 with a slope of 54.2 +/- 0.4 mV pH(- 1) at 20 +/- 1 degreesC. It has a rapid potential-response to changes of pH, high ion selectivity towards lithium, sodium and potassium, and other characteristics comparable to those reported for the conventional pH glass membrane electrode. It appears to be a suitable potentiometric indicator electrode specifically for hydrofluoric acid solutions. (C) 2003 Elsevier B.V. All rights reserved.

Açıklama

Anahtar Kelimeler

p-tert-butylcalix[4]arene-oxacrown-4, hydrogen ion-selective, PVC membrane electrode, potentiometry

Kaynak

TALANTA

WoS Q Değeri

Q1

Scopus Q Değeri

Q1

Cilt

62

Sayı

1

Künye